Eddy current modules can measure non-conductive coatings on non-ferrous metals including paint, anodising, vitreous enamel, plastic, epoxy, varnish power coating, glass fibre, rubber etc. on aluminium, copper, brass, austentitic (non-magnetic) stainless steel, and other similar materials.
Operating temperature: 0 - 50°C (32 - 120°F).
Minimum substrate thickness:
Ferrous*: 0.3 mm (12 mils).
Non ferrous: 0.1 mm (4 mils).
(*with 2 point calibration 1 mm for ±1 % accuracy over probe range).
Measuring range: 0 - 1500 µm.
Resolution steps:
0,1 µm (0 - 200 µm).
0,05 µm (200 - 500 µm).
1,0 µm (500 - 1500 µm).
Accuracy: ±1 % or ±1 µm.
Minimum convex surface diameter: 44 mm (1.73").
Minimum concave surface radius: 25 mm (0.98").
Headroom: 90 mm (3.54").
Minimum sample diameter: 8 mm (0.31").
Probe calibration point: 250 µm.
Operating temperature: 0 - 50°C (32 - 120°F).
Minimum substrate thickness:
Ferrous*: 0.3 mm (12 mils).
Non ferrous: 0.1 mm (4 mils).
(*with 2 point calibration 1 mm for ±1 % accuracy over probe range).
Measuring range: 0 - 1500 µm.
Resolution steps:
0,1 µm (0 - 200 µm).
0,05 µm (200 - 500 µm).
1,0 µm (500 - 1500 µm).
Accuracy: ±1 % or ±1 µm.
Minimum convex surface diameter: 44 mm (1.73").
Minimum concave surface radius: 25 mm (0.98").
Headroom: 90 mm (3.54").
Minimum sample diameter: 8 mm (0.31").
Probe calibration point: 250 µm.